SAXS Transmission Calculator
Thickness for a target transmission
–
Transmission of a given thickness
–

Transmission vs. thickness

Target Your thickness Optimum 1/µ

Measurement setup

Click the source, sample or transmitted beam to change it
scattered X-rays (SAXS) of optimum (at T ≈ 37 %) – X-ray source – incident beam I₀ holder walls sample – – beamstop 2D detector
The beam passes the holder walls and a sample of thickness t; a fraction T = e−µt · Tholder comes out and is caught by the beamstop, the rest is absorbed or scattered. A thicker sample scatters more but absorbs more, so the SAXS signal ∝ t·e−µt is largest at t = 1/µ (T ≈ 37 %).

How it works

The linear attenuation coefficient of the sample is built up from its phases:

µ = (1 − P) · Σi φi ρi Σj wij (µ/ρ)j(E)

φi is the volume fraction of phase i in the solid (converted from wt % if needed), ρi its density, wij the mass fraction of element j in it, and P the porosity. Transmission follows Beer–Lambert, T = exp(−µt), and the thickness for a target transmission is t = −ln(T)/µ.

Optimal SAXS thickness. The scattered intensity scales as t·e−µt, which peaks at t = 1/µ, i.e. T ≈ 37 %.

Data. Total cross sections (photoabsorption + coherent + incoherent scattering) from Elam, Ravel & Sieber, Radiat. Phys. Chem. 63 (2002) 121, interpolated with log-log cubic splines.

This is an estimate. Real densities, residual solvent, and surface roughness all shift the answer, so measure the transmission on the instrument.

Density from unit cell

Crystal density from the lattice parameters: ρ = Z·M / (NA·V).

–